    <title locale="en_US">Filters Effect on Image Quality of 177Lu-DOTATATE SPECT Images for Neuro-Endocrine Tumors Scanning</title>
    <FirstPage>1</FirstPage>
    <LastPage>8</LastPage>
    <AuthorList>
      <Author>
        <FirstName>Hussein</FirstName>
        <LastName>Dakhil</LastName>
        <affiliation locale="en_US">Biomedical Engineering and Medical Physics Department, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Mohammad Reza</FirstName>
        <LastName>Deevband</LastName>
        <affiliation locale="en_US">Department of Bioengineering and Medical Physics, Shahid-Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Mahasti</FirstName>
        <LastName>Amoui</LastName>
        <affiliation locale="en_US">Department of Bioengineering and Medical Physics, Shahid-Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Ahmad</FirstName>
        <LastName>Mostaar</LastName>
        <affiliation locale="en_US">Department of Bioengineering and Medical Physics, Shahid-Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Mahdi</FirstName>
        <LastName>Ghorbani</LastName>
        <affiliation locale="en_US">Biomedical Engineering and Medical Physics Department, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Mohsen</FirstName>
        <LastName>Khosroabadi</LastName>
        <affiliation locale="en_US">Department of Medical Physics, North Khorasan University of Medical Sciences, Bojnurd, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Mohammad Ali</FirstName>
        <LastName>Ghodsirad</LastName>
        <affiliation locale="en_US">Nuclear Medicine Department, Shohada-e-Tajrish Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
    </AuthorList>
    <History>
      <PubDate PubStatus="received">
        <Year>2021</Year>
        <Month>10</Month>
        <Day>26</Day>
      </PubDate>
      <PubDate PubStatus="accepted">
        <Year>2023</Year>
        <Month>07</Month>
        <Day>29</Day>
      </PubDate>
    </History>
    <abstract locale="en_US">Purpose: The quality of the image in SPECT scans depends on the imaging parameters which are determined experimentally in the field of nuclear medicine. Designing a dedicated scanning parameter for 177Lu-DOTATATE SPECT images is required to optimize reconstruction. Therefore, this study aims to evaluate the effect of different filters on image quality for bone SPECT scanning using 177Lu-DOTATATE.
&#xD;

Materials and Methods: The filtered back-projection reconstruction method was used in neuro-endocrine tumor scanning using 177Lu-DOTATATE. Three hours after injection of 177Lu-DOTATATE, SPECT scans from 177Lu-DOTATATE for 30 patients were acquired using a dual-head EvoExel detector system. Seven parameters were considered, including the contrast/noise ratio, injection activity, uptake duration, acquisition time per injection, frame time, measuring time, and type of filters.
&#xD;

Results: In all cases, the application of different filters increased Contrast to Noise Ratio (CNR) (9.1%, 1.8%, 10.9%, 61.8%, 23.6%, 29.1%, and 58.2% for Wiener, Butterworth, Parzen, Metz, Ramp, Shepp-Logan, and Hamming filters, respectively). The percentage of increase in Signal-to-Noise Ratio (SNR) was 3.3%, 1.7%, -24%, 21.7%, 9.8%, 11.9%, and 20.6%.
&#xD;

Conclusion: Based on the quantitative analysis of the results, the application of the Metz filter (power 2) and Hamming filter (with 0.27, 0.47, and 0.67 cut-off frequencies) on SPECT scans of neuro-endocrine tumors is recommended because of their capacity to provide high-quality images.</abstract>