    <title locale="en_US">Evaluation of Setup Errors in Two Radiotherapy Centers from a Physicist&#x2019;s Point of View</title>
    <FirstPage>155</FirstPage>
    <LastPage>162</LastPage>
    <Language>EN</Language>
    <AuthorList>
      <Author>
        <FirstName>Keyvan</FirstName>
        <LastName>Jabbari</LastName>
        <affiliation locale="en_US">Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran.</affiliation>
      </Author>
      <Author>
        <FirstName>Nima</FirstName>
        <LastName>Rostampour</LastName>
        <affiliation locale="en_US">Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran.</affiliation>
      </Author>
    </AuthorList>
    <History>
      <PubDate PubStatus="received">
        <Year>2015</Year>
        <Month>12</Month>
        <Day>27</Day>
      </PubDate>
      <PubDate PubStatus="accepted">
        <Year>2015</Year>
        <Month>12</Month>
        <Day>27</Day>
      </PubDate>
    </History>
    <abstract locale="en_US">Purpose- The setup verification is one of the important issues in radiation therapy.&#xA0;In this study, an experienced physicist evaluated the setup of the patients during&#xA0;treatments where many unexpected errors were observed.

Methods- The physicist spent few hours a week in the treatment room and recorded&#xA0;any error that happened in the treatment setup from the physicist&#x2019;s point of view.&#xA0;In some errors, a follow up dosimetry was performed to evaluate the effect of a&#xA0;specific error.

Results- The errors were divided into a few categories. Out of 1000 patients in a 3&#xA0;years period, various minor and major errors were observed for 115 patients. Most&#xA0;of the errors were in treatment field&#x2019;s shape and size. There were also few mistakes&#xA0;made by technicians due to the lack of conceptual understanding particularly when&#xA0;the electron shield was placed too far from the skin. Results of film dosimetry&#xA0;revealed that this can cause a severe underdose of the tumor and an overdose of&#xA0;the shielded area.

Conclusion- Many of the recorded setup errors in this study were related to the
setup protocols for a particular center. However, there are many mistakes such
as mistakes of the technicians and physicists that can be prevented with proper
trainings.</abstract>