    <title locale="en_US">Estimation of Dose Perturbation Due to the Presence of Metal Hip Prostheses in Radiotherapy with Electron and Photon Beams: A Monte Carlo Study</title>
    <FirstPage>642</FirstPage>
    <LastPage>650</LastPage>
    <AuthorList>
      <Author>
        <FirstName>Morteza</FirstName>
        <LastName>Hashemizadeh</LastName>
        <affiliation locale="en_US">Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.</affiliation>
      </Author>
      <Author>
        <FirstName>Mansour</FirstName>
        <LastName>Zabihzadeh</LastName>
        <affiliation locale="en_US">Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.</affiliation>
      </Author>
      <Author>
        <FirstName>Omid</FirstName>
        <LastName>Azadbakht</LastName>
        <affiliation locale="en_US">Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.</affiliation>
      </Author>
      <Author>
        <FirstName>Masoud</FirstName>
        <LastName>Jamali</LastName>
        <affiliation locale="en_US">Radiation Oncologist, Technical Manager in Sarerm Hospital, Tehran, Iran</affiliation>
      </Author>
    </AuthorList>
    <History>
      <PubDate PubStatus="received">
        <Year>2023</Year>
        <Month>11</Month>
        <Day>06</Day>
      </PubDate>
      <PubDate PubStatus="accepted">
        <Year>2023</Year>
        <Month>12</Month>
        <Day>22</Day>
      </PubDate>
    </History>
    <abstract locale="en_US">Purpose: The impact of various hip prosthesis materials on the amount of dose perturbation generated by 9 MeV electron and 18 MV photon beams during pelvic radiation therapy is analyzed in this research.
&#xD;

Materials and Methods: The Varian 2100 C/D LINAC head for the electron (9 MeV) and photon (18 MV) modes and a water phantom with a realistic hip prosthesis were modeled using the Monte Carlo (MC) code; MCNPX (Ver. 2.6.0) and were benchmarked for measurement. Four different materials, including Cobalt&#x2013;Chromium&#x2013;Molybdenum-alloy (CCM), Stainless Steel (SS), Titanium, and Titanium-alloy (Ti-alloy) were evaluated. The changes in electron and photon fluences and dose perturbations due to the presence of the hip prostheses were investigated.
&#xD;

Results: An increased dose of 13.29%, 13.77%, 6.16%, and 5.93% for 9 MeV and 30.43%, 33.05%, 10.89%, and 11.27% for 18 MV was calculated for Ti-alloy, Ti, CCM, and SS prosthesis, respectively. At 0.5 mm distance from the prosthesis, the Electron Backscatter Factor (EBF) of 1.31, 1.33, 1.15, and 1.14 for 9 MeV and 1.32, 1.34, 1.11, and 1.12 for 18 MV was calculated for Ti-alloy, Ti, CCM, and SS prosthesis, respectively. Dose perturbation is higher at a near distance from the prosthesis; by reducing the distance from the Ti-alloy prosthesis (1.5 to 0.5 mm), an increased dose of 7.70% and 5.54% resulted in 18 MV and 9 MeV, respectively. The dose decreases of up to 21% behind the hip prosthesis were calculated for 18 MV.
&#xD;

Conclusion: It is essential to consider the dose perturbation due to the presence of a hip prosthesis to achieve the optimal treatment planning in radiotherapy.</abstract>