Purpose: Gamma Knife is applied as a superseded tool for inaccessible lesions surgery delivering a single high dose to a well-defined target through 201 small beams. Monte Carlo simulations can be an appropriate supplementary tool to determine dosimetric parameters in small fields due to the related dosimetry hardships.
Materials and Methods: EGSnrc/BEAMnrc Monte Carlo code was implemented to model Gamma Knife 4C. Single channel geometry comprising stationary and helmet collimators was simulated. A point source was considered as a cylindrical Cobalt source based on the simplified source channel mode. All of the 201 source channels were arranged in spherical coordinate by EGSnrc/DOSXYZnrc code to calculate dose profiles. The simulated profiles at the isocentre point in a spherical head phantom 160 mm in diameter along three axes for 4, 8, 14, and 18 mm field sizes were compared to those obtained by another work using MCNP code.
Results: Based on the results, the BEAMnrc and MCNP dose profiles matched well apart from the 18 mm profiles along X and Y directions with the average gamma index of 1.36 and 1.18, respectively. BEAMnrc profiles for 14 and 18 mm field sizes along X and Y axes were entirely flat in plateau region, whereas MCNP profiles represented variations as well as round shape. Besides, considering the identical results, radioactive source can be modeled by a point source instead of cylindrical one.
Conclusion: Thus, the EGSnrc/BEAMnrc code is recommended to simulate Gamma Knife machine as it is regarded as the most accurate computer program to simulate photon and electron interactions.

1- Cheung JY, Yu KN, Yu CP, et al., “Monte Carlo calculation of single‐beam dose profiles used in a gamma knife treatment planning system” Med. Phys, vol. 26, no. 7, pp. 1252-6, 1998.
2- Kim IH, Lim DH, Kim S, Hong S, Kim BK, Kang WS, Wu HG, Ha SW, Park CI, “Extracranial doses during stereotactic radiosurgery and fractionated stereotactic radiotherapy measured with thermoluminescent dosimeter in vivo”. P-7-66. From http://www. irpa. net/irpa10/cdrom/00684. pdf; 2000 May 1.
3- Moskvin V, DesRosiers C, Papiez L, et al., “Monte Carlo simulation of the Leksell Gamma Knife®: I. Source modelling and calculations in homogeneous media” Phys. Med. Biol, vol. 47, no. 12, pp. 1995-2011, 2002.
4- Al-Dweri FM, Lallena AM, “A simplified model of the source channel of the Leksell Gamma Knife®: testing multisource configurations with PENELOPE” Phys. Med. Biol, vol. 49, no. 15, pp. 3441-53, 2004.
5- Xiong W, Huang D, Lee L, et al., editors, “Implementation of Monte Carlo simulations for the Gamma Knife system” J. Phys: Conference Series:pp. 1-10, 2007.
6- Romano F, Sabini M, Cuttone G, et al., “Geant4-based Monte Carlo Simulation of the Leksell Gamma Knife®” Nuclear Science Symposium Conference Record; 26 Oct.-3 Nov. 2007; Honolulu, HI, USA: IEEE;. P. 2581-6, 2007.
7- Battistoni G, Cappucci F, Bertolino N, et al., “FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion radiosurgery system: homogeneous media” Physica Medica, vol. 29, no. 6, pp. 656-61, 2013.
8- Cheung JY, Yu KN, Ho RT, Yu CP, “Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit” Medical physics, vol. 26, no. 7, pp. 1252-6, 1999.
9- Moskvin V, Timmerman R, DesRosiers C, Randall M, DesRosiers P, Dittmer P, Papiez L, “Monte Carlo simulation of the Leksell Gamma Knife®: II. Effects of heterogeneous versus homogeneous media for stereotactic radiosurgery” Physics in Medicine & Biology, vol. 49, no. 21, pp. 4879, 2004.
10- Cheung JY, Yu KN, “Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo code” Medical physics, vol. 33, no. 1, pp. 41-5, 2006.
11- Trnka J, Novotny Jr J, Kluson J, “MCNP‐based computational model for the Leksell Gamma Knife” Medical physics, vol. 34, no. 1, pp. 63-75, 2007.
12- Gholami S, Kamali Asl A, Aghamiri SM, Allahverdi M, “Gamma Knife Simulation Using the MCNP4C Code and the Zubal Phantom and Comparison with Experimental Data” Iranian Journal of Medical Physics, vol. 7, no. 2, pp. 57-63, 2010.
13- Mahmoudi A, Shirazi A, Geraily G, Bakhshi M, Maleki M, “Penumbra width determination of single beam and 201 beams of Gamma Knife machine model 4C using Monte Carlo simulation” Journal of Radiotherapy in Practice, vol. 18, no. 1, pp. 82-7, 2019.
14- Mahmoudi A, Geraily G, Hadisinia T, Shirazi A, Najafzadeh M, “Beam penumbra reduction of Gamma Knife machine model 4C using Monte Carlo simulation” Computer Methods and Programs in Biomedicine, 2020 May 1;188:105261.
15- Rogers D, Walters B, Kawrakow I 2011 “DOSXYZnrc users manual”. Nrc Report Pirs. Ottawa, Canada. National Research Council of Canada; 2017.
16- Rogers D, Walters B, Kawrakow I. “BEAMnrc users manual”. Nrc Report Pirs. Ottawa, Canada. National Research Council of Canada; 2017.
17- McGowan H, Faddegon B, Ma C , “STATDOSE for 3D dose distributions. NRCC report PIRS-0509 (F) “. Nrc Report Pirs, Ottawa, Canada: NRCC, 2007.
Files
IssueVol 7 No 1 (2020) QRcode
SectionOriginal Article(s)
DOI https://doi.org/10.18502/fbt.v7i1.2723
Keywords
Gamma Knife Monte Carlo N-Particle Point Source Electron Gamma Shower National Research Council /BEAM National Research Council

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Mahmoudi A, Geraily G. EGSnrc/BEAMnrc-Based Monte Carlo Simulation of the Gamma Knife 4C versus MCNP Code in Homogeneous Media. Frontiers Biomed Technol. 2020;7(1):33-40.