Assessment of CT Imaging Protocols Impacts on Calculation of Point Dose in Water Phantom using Radiotherapy Treatment Planning System

Abstract

Purpose: Point dose calculation in the Treatment Planning System (TPS) is performed using Computed Tomography (CT) images because CT images data have the tissue electron density information. The effect of CT imaging protocols on the calculation of point doses in TPS is one of the most important subjects that was evaluated in this study.
Materials and Methods: CT scan imaging was performed from cylindrical water phantom using three scanner systems and different imaging technical parameters. The CT images data were irradiated in TPS to delivering a 200 cGy radiation dose to the center of the phantom with 6 and 15MV X-Ray photon energy with multiple radiation fields and Monitor Unit (MU) were separately calculated. In the TPS, a virtual water phantom with the same characteristic as CT image phantom was simulated and irradiated with similar conditions. The difference in MU values obtained from two irradiation methods in TPS was compared with Wilcoxon nonparametric test.  
Results: Variations of mA, kV, Pitch, slice thickness, and kernel as CT imaging parameters have not significantly affected radiotherapy point dose calculation (<2%). CT imaging protocols as a thin slice, 80 kV, and sharp kernel have the greatest difference between CT image-based calculation and designed phantom calculation in TPS where wedge field and 6 MV photon energy were used.
Conclusion: The use of CT images obtained with multiple protocols can be used without having a significant effect on the dose calculations of the treatment planning system.

1- Parker RP, Hobday PA, Cassell KJ. "The direct use of CT numbers in radiotherapy dosage calculations for inhomogeneous media," Phys. Med. Biol; vol. 24, pp.802–9, 1979.
2- WitoldSkrzyn ´ski, SylwiaZielin ´ska-Da˛browska, Marta Wachowicz, Wioletta S ´lusarczykKacprzyk,Paweł F. ukołowicz, WojciechBulski. "Computed Tomography as a Source of Electron DensityInformation for Radiation Treatment Planning,"
Strahlenther Onkol, vol. 186, pp. 327–33, 2010.
3- Seco J, Evans PM. "Assessing the effect of electron density in photon dosecalculations," Med Phys, vol. 33,
pp. 540–52, 2006.
4- Knöös T, Nilsson M, Ahlgren L. "A method for conversion of Hounsfield number toelectron density and 5- Georg D, Georg P, Hillbrand M, et al. "Assessment of improved organ atrisk sparing for advanced cervix carcinoma utilizing precision radiotherapytechniques,"StrahlentherOnkol, vol.184, pp. 586–91, 2008.
6- Goerlitz E, Albers D, Cremers F, Schmidt R. "Influence of the CT calibration process on the dose computation in radiotherapy treatment planning,"Proceedings of the International Conference on Quality Assurance and NewTechniques in Radiation Medicine, Vienna, November 13–15, pp. 226–7, 2006.
7- Liu RR, Prado KL, Cody D. "Optimalacquisition parameter selection for CT simulators in radiation oncology," J ApplClin MedPhys, vol. 9, pp. 151–60, 2008.
8- Kearns D, McJury M. "Commissioning of a new CT simulator I: CT simulator hardware," J RadiotherPract, vol.6, pp. 153–62, 2007.
9- Ebert MA, Lambert J, Greer PB. "CT-ED conversion on a GE Lightspeed-RT scanner: the influence of scanner settings," AustralasPhysEngSci Med, vol. 31, pp. 154–9, 2008.
10- Davis AT, Palmer AL, Nisbet A. "Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review," Br J Radiol, vol. 90(1076), pp. 20160406, 2017.
11- Zurl B, Tiefling R, Winkler P, Kindl P, Kapp K. "Hounsfield units variations: the impact on CT-density based conversion tables and their effect on dose distribution," StrahlentherOnkol, vol. 190, pp. 88–93, 2014.
12- Rutonjski L, Petrovic P, Baucal M, Teodorovic M, Cudic O, Gershkevitsh E, et al. "Dosimetric verification of radiotherapy treatment planning systems in Serbia: national audit," RadiatOncol, vol. 7, pp. 155, 2012
13- Mahur M, Gurjar O P, GROVER RK, Negi PS, Sharma R, Singh A, Singh M. "Evaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System," IRANIAN JOURNAL OF MEDICAL PHYSICS, vol. 14, pp. 149-154, 2017
Files
IssueVol 7 No 2 (2020) QRcode
SectionOriginal Article(s)
DOI https://doi.org/10.18502/fbt.v7i2.3857
Keywords
Computed Tomography Scan Imaging Protocol Point Dose Radiotherapy

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Fallah Mohammadi G, Mihandoost E. Assessment of CT Imaging Protocols Impacts on Calculation of Point Dose in Water Phantom using Radiotherapy Treatment Planning System. Frontiers Biomed Technol. 2020;7(2):112-117.