Original Article

Evaluation of DAP Values Obtained from Chest X-rays in Children under 12 Years of Age Referred to Educational Hospitals of Birjand University of Medical Sciences in 2020

Abstract

Purpose: Dose Area Product (DAP) is a quantity for radiation risk assessment in diagnostic X-ray tests. Children's tissues are up to 10 times more sensitive to radiation than adults, and life expectancy is higher in children than adults, as well as a higher risk of hematopoietic and mass malignancies in them. Therefore, this study aimed to measure DAP values for X-ray fields adjusted by Birjand radio-technologists in chest X-rays of children under 12 years of age.

Materials and Methods: 233 children from Birjand University hospitals who performed chest X-rays were included in the study. To collect data related to DAP, the DAP meter model KermaX plus SDP was used. It should be noted that no intervention was performed in the patient's imaging method and at the time of radiation and measurement of the DAP values, there was no need for the patient's presence. In the end, the measured DAP values were compared with DAP values of other studies. Data were analyzed using SPSS software version 22 at 5% error level using Anova, t-test, and Pearson correlation tests.

Results: Out of 233 patients who were included in the study, 134 males (57.5%) and 99 females (42.5%) participated in the study, it should be noted that there was no significant difference between the mean of DAP in male and female (p=0.52). In our study, the average DAP was 5.78 ± 3.54 μGy.m2 and DAP values in the range of 0.55 μGy.m2 to 15.54 μGy.m2 that were higher than the average of other studies and there was a significant difference. There was a direct relationship between radiation field dimensions and DAP values so that as the dimensions of the radiation field increase, the DAP value increases. There was a significant difference between the mean DAP of the lowest and the highest age groups, lowest and the highest age groups, and lowest and the highest height groups of patients.

Conclusion: In our study, it was observed that there is a significant relationship between patients' weight, age, and height, radiation field dimensions with DAP values. The amount of DAP in the present study was significantly higher than in other studies. The most important effective parameter in DAP is the radiation field size and if sufficient optimization is done in imaging parameters (kVp, mAS, field size), the DAP values will be greatly reduced.

1- Hall, Eric J., and Amato J. Giaccia. "Radiobiology for the Radiologist.", Vol. 6. (2006).
2- Okeji, M. C., Anakwue, A. M., & Agwuna, K. "Radiation exposure from diagnostic radiography: an assessment of X-ray beam collimation practice in some Nigerian Hospitals.", Internet Journal of Medical Update-EJOURNAL. 5(2),‏ (2010).
3- Bouzarjomehri F. "Patient dose in routine Xray examinations in Yazd state.", Iran J RadiatRes; 1(4): 199-204, (2004).
4- Johnston D A, Brennan P C. "Reference dose levels for patients undergoing commondiagnostic X-ray examinations in Irish hospitals.", Br J Radiol; 73: 396 – 402, (2000).
5- Dose Area Product available at https://en.wikipedia.org/wiki/Dose_area_product. Accessed in 5/15/2019.
6- "National Radiological Protection Board (NRPB) report R263, 1993.", Radiation Exposure of the UK population– (2000).
7- Zagoria RJ, Dixon RL. "Radiology of urolithiasis: implications of radiation exposure and new imaging modalities.", Adv Chronic Kidney Dis; 16(1): 48-51. (2009).
8- Brenner D, Elliston C, Hall E, Berdon W. "Estimated risks of radiation-induced fatal cancer from pediatric CT." AJR Am J Roentgenol; 176(2): 289-96, (2001).
9- Eeg KR, Khoury AE, Halachmi S, Braga LH, Farhat WA, Bagli DJ, et al. "Single center experience with application of the ALARA concept to serial imaging studies after blunt ren trauma in children--is ultrasound enough?", J Urol; 181(4): 1834-40, (2009).
10- "United Nation Scientific Committee on the effects of Atomic Radiation. Sources and effects of ionizing radiation.", Report to the general assembly with scientific annexes, (2000).
11- Commission of the European Communities. Quality criteria for diagnostic images and patient exposure trial. Document XII/268/90. Brussels: CEC 1990.
12- Kisielewicz, K., Truszkiewicz, A., Wach, S., & Wasilewska–Radwańska, M. "Evaluation of dose area product vs. patient dose in diagnostic X-ray units.", Physica Medica, 27(2), 117-120, (2011).
13- Fauber, T. L., & Dempsey, M. C. "X-ray field Size and patient dosimetry.", Radiologic technology. 85(2), 155-161, (2013).
14- Karami, V., Zabihzadeh, M., Gilavand, A., & Shams, N. "Survey of the use of X-ray beam collimator and shielding tools during infant chest radiography.", International Journal of Pediatrics., 4(4), 1637-1642, (2016).
15- Nnamdi N. Jibiri and Christopher J. Olowookere. "Evaluation of dose-area product of common radiographic examinations towards establishing a preliminary diagnostic reference levels (PDRLs) in Southwestern Nigeria." J Appl Clin Med Phys, 17(6):392-404, (2016).
16- Billinger, J., Nowotny, R., & Homolka, P. "Diagnostic reference levels in pediatric radiology in Austria.", European radiology, 20(7), 1572-1579, (2010).
17- Roch, P., & Aubert, B. "French diagnostic reference levels in diagnostic radiology, computed tomography and nuclear medicine: 2004–2008 review.", Radiation protection dosimetry, 154(1), 52-75, (2013).
18- McDonald, S., Martin, C. J., Darragh, C. L., & Graham, D. T. "Dose–area product measurements in paediatric radiography.", The British journal of radiology, 69(820), 318-325,‏ (1996).
19- Chida K, Ohno T, Kakizaki S, Takegawa M, Yuuki H, Nakada M, Takahashi S, Zuguchi M. "Radiation dose to the pediatric cardiac catheterization and intervention patient.", American Journal of Roentgenology;195(5):1175-9, (2010).
Files
IssueVol 9 No 3 (2022) QRcode
SectionOriginal Article(s)
DOI https://doi.org/10.18502/fbt.v9i3.9644
Keywords
Dose Area Product Pediatric Radiography Chest X-Ray Radiation Risk

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Ghayour-Saffar N, Ehsanbakhsh A, Keshtkar M, Pandesh S. Evaluation of DAP Values Obtained from Chest X-rays in Children under 12 Years of Age Referred to Educational Hospitals of Birjand University of Medical Sciences in 2020. Frontiers Biomed Technol. 2022;9(3):185-190.