Original Article

Assessment of SPECT Image Reconstruction in Liver Scanning Using 99mTc/ EDDA/ HYNIC-TOC

Abstract

Introduction: SPECT is reconstructed using iterative techniques incorporating photon attenuation correction based on the X ray transmission map and scatter correction. Since the quality of image in SPECT scans depends on the imaging parameters that are determined experimentally in the field of nuclear medicine, designing a dedicated scanning method for 99mTc/SPECT is need to improve the accuracy of disease diagnostic. Therefore, in this study with the aims of assessment and evaluation the effect of different filters on image quality, assessment of quality criteria on image before and after optimization and determine optimum algorithm for reconstruction on liver scanning using 99m Tc for SPECT studies.
Material and Method: Filtered Back-projection reconstruction method had been used in liver scanning using 99m Tc/EDDA/HYNIC-TOC for SPECT image and effect of different filters on SPECT imaging have been evaluated.
Results and Conclusion: Based on results, we suggest the Hamming filter to be used for visual analysis of Liver SPECT because of its ability to produce the high-quality image. Instead, the Butterworth filter is suggested for quantitative analysis because of its ability to balance between image quality and noise.

1- Genes SG. "Role of the liver in hormone metabolism and in the regulation of their content in the blood." Arkh Patol; 39(6): 74-80, (1977).
2- Espersen C, Borgwardt L, Larsen PN, Andersen TB, Stenholt L, Petersen LJ. "Nuclear imaging methods for the prediction of postoperative morbidity and mortality in patients undergoing localized, liver-directed treatments: a systematic review." EJNMMI Res; 10(1): 101, (2020).
3- Lyra M, Ploussi A. "Filtering in SPECT Image Reconstruction." Int J Biomed Imaging 2011; 693795, (2011).
4- Gullberg GT, Reutter BW, Sitek A, Maltz JS, Budinger TF. "Dynamic single photon emission computed tomography-basic principles and cardiac applications." Phys Med Biol; 55(20): 111-91, (2010)
5- Ljungberg M, Pretorius PH. "SPECT/CT: an update on technological developments and clinical applications." Br J Radiol; 91(1081): 20160402, (2018).
6- Velikyan I. "Prospective of ⁶⁸Ga-radiopharmaceutical development." Theranostics; 4(1): 47-80, (2013).
7- Sayed IS, Ismail SS. Comparison of Low-Pass Filters for SPECT Imaging. Int J Biomed Imaging; 2020: 9239753, (2020).
8- Frey EC, Humm JL, Ljungberg M. "Accuracy and precision of radioactivity quantification in nuclear medicine images." Semin Nucl Med; 42(3): 208-18, (2012).
9- Saib DMA, Azman NZN, Said MA, Aseri MIM, Almarri HM, Ramli RM. "Evaluation of butterworth post-filtering effects on contrast and signal noise to ratio values for SPECT images reconstruction." Radiat Phys Chem; 192: 109932, (2022).
10- Park CR, Kang S-H, Lee Y. "Median modified wiener filter for improving the image quality of gamma camera images." Nucl Eng Technol; 52(10): 2328-33, (2020).
11- Buvat I, Frey E, Green A, Ljungberg M. "IAEA human health report number 9. Quantitative nuclear medicine imaging: concepts, requirements and methods.", IAEA, Vienna, (2014).
12- Dorbala S, Ananthasubramaniam K, Armstrong IS, Chareonthaitawee P, DePuey EG, Einstein AJ, et al. "Single Photon Emission Computed Tomography (SPECT) Myocardial Perfusion Imaging Guidelines: Instrumentation, Acquisition, Processing, and Interpretation." J Nucl Cardiol; 25(5): 1784-846, (2018).
13- Al Moudi M, Sun Z-H. "Diagnostic value of 18F-FDG PET in the assessment of myocardial viability in coronary artery disease: A comparative study with 99mTc SPECT and echocardiography." J Geriatr Cardiol; 11(3): 229-236, (2014).
14- Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. "Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction." J Nucl Med; 51(6): 921-8, (2010).
15- Lyra M, Ploussi A, Rouchota M, Synefia S. "Filters in 2D and 3D cardiac SPECT image processing." Cardiol Res Pract 2014; (2014).
16- Mohseni S, Kamali-Asl A, Bitarafan-Rajabi A, Entezarmahdi SM, Shahpouri Z, Yaghoobi N. "Effects of filtration on right ventricular function by the gated blood pool SPECT." Ann Nucl Med; 29(4): 384-90, (2015).
17- Davidsson A, Olsson E, Engvall J, Gustafsson A. "Influence of reconstruction algorithms on image quality in SPECT myocardial perfusion imaging." Clin Physiol Funct Imaging; 37(6): 655-62, (2017).
18- Choi MS, Kang HG. "Transient noise reduction in speech signal with a modified long-term predictor." EURASIP J Adv Signal Process; 2011(1): 141, (2011).
19- King MA, Schwinger RB, Doherty PW, Penney BC. "Two-dimensional filtering of SPECT images using the Metz and Wiener filters." J Nucl Med; 25(11): 1234-40, (1984).
20- Pandey A, Pant G, Malhotra A. "Standardization of SPECT filter parameters." Indian J Nucl Med; 19(2): 30-5, (2004).
21-Shahbazi-Gahrouei D, Arabpour A, Kermani S, Rastgoo F. "Assessment of gated single photon emission computerized tomography cardiac wall motion by using different reconstruction methods and filters in comparison with quantitative coronary angiography." J Med Sci; 8(4): 342-9, (2008).
Files
IssueVol 11 No 2 (2024) QRcode
SectionOriginal Article(s)
DOI https://doi.org/10.18502/fbt.v11i2.15342
Keywords
Single Photon Emission Computed Tomography Imaging Filtering Filter Design Image Quality Image Reconstruction

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Shareef Abdulhusein N, Deevband MR, Ghodsirad MA, Behmadi M, Mehri-Kakavand G. Assessment of SPECT Image Reconstruction in Liver Scanning Using 99mTc/ EDDA/ HYNIC-TOC. Frontiers Biomed Technol. 2024;11(2):247-254.