Original Article

Quantitative Nail Fold Capillary Blood Flow Using Capillaroscopy System and ImageJ Software in Healthy Individuals

Abstract

Purpose: Measuring the blood flow velocity in capillaries is a useful method for diagnosing many diseases. Despite increasing interest in nailfold capillaroscopy, objective measures of capillary structure and blood flow have been rarely studied. This study aimed to measure the blood flow velocity along the capillary central line using capillaroscopy system, and also ImageJ software used Scale-Invariant Feature Transform (SIFT) tracking algorithms and Kalman filter for image processing.

Materials and Methods: The Red Blood Cells (RBCs) velocity in capillaries of finger nailfold was measured in 12 normal cases using a novel capillaroscopy system. The measurements of the velocity were performed at 12 points in nailfold regions by two observers separately. The image processing and automated measurement take 1-2 min per nailfold. FFmpeg software was used to convert the images format, and then the images were imported to ImageJ software and segmented. SIFT tracking algorithms and Kalman filter were used to filter noise and irregularities in the images. For reproducibility, the velocity distribution values obtained by the two performers, and Paired T-Test was used. The reliability of a measuring instrument or calculation method depends on the tools obtained using Cronbach's alpha. To assess the repeatability of the algorithm, the capillary velocity values were executed at different times with 24-hour intervals using a coefficient of variance method.

Results: The calculated RBCs velocity was in the range of 0.05-0.16 mm/s. The results based on Cronbach's alpha analysis for reliability factor was 0.97, with a good correlation among the measurements, 0.85. The average velocity (along with standard deviation) for repeatability at three different times was obtained 0.1195 ± 0.0246, 0.0974 ± 0.0221, and 0.0962 ± 0.0202 mm/s, demonstrating that there was no statistically variation between these measurements (P-value > 0.05). The velocity results for the two observers were 0.811 ± 0.392 and 0.819 ± 0.325 mm/s, indicating a good reproducibility between them (P-value = 0.959).

Conclusion: For the measurements of nailfold capillaries velocity, there was good/reasonable reliability, repeatability, and reproducibility. The results indicated a good accuracy of capillaroscopy system and ImageJ software with SIFT algorithm and Kalman filter, which can be used as an appropriate tool for determining the rate of nailfold blood flow velocity.

1- Gurov I, Volkov M, Margaryants N, Pimenov A, Potemkin A. "High-speed video capillaroscopy method for imaging and evaluation of moving red blood cells.", Opt Lasers Eng;104:244–51, (2018).
2- Smith V, Thevissen K, Trombetta AC, Pizzorni C, Ruaro B, Piette Y, et al. "Nailfold capillaroscopy and clinical applications in systemic sclerosis.", Microcirculation.; 23(5):364–72, (2016).
3- Allen J, Howell K. "Microvascular imaging: techniques and opportunities for clinical physiological measurements.", Physiol Meas.;35(7):R91, (2014).
4- Daly SM, Leahy MJ. "‘Go with the flow’: a review of methods and advancements in blood flow imaging.", J Biophotonics;6(3):217–55. (2013).
5- Shih T-C, Zhang G, Wu C-C, Hsiao H-D, Wu T-H, Lin K-P, et al. "Hemodynamic analysis of capillary in finger nail-fold using computational fluid dynamics and image estimation.", Microvasc Res;81(1):68–72, (2011).
6- Mugii N, Hasegawa M, Hamaguchi Y, Tanaka C, Kaji K, Komura K, et al. "Reduced red blood cell velocity in nail-fold capillaries as a sensitive and specific indicator of microcirculation injury in systemic sclerosis." Rheumatology; 48(6):696–703, (2009).
7- Cutolo M, Pizzorni C, Sulli A. "Capillaroscopy." Best Pract Res Clin Rheumatol;19(3):437–52, (2005).
8- Watanabe M, Matsubara M, Sanada T, Kuroda H, Iribe M, Furue M. "High speed digital video capillaroscopy: nailfold capillary shape analysis and red blood cell velocity measurement." J Biomech Sci Eng;2(2):81–92, (2007).
9- Gurfinkel YI, Suchkova OV, Sasonko ML, Priezzhev AV. "Implementation of digital optical capillaroscopy for quantifying and estimating the microvascular abnormalities in type 2 diabetes mellitus.", In: Saratov Fall Meeting 2015: Third International Symposium on Optics and Biophotonics and Seventh Finnish-Russian Photonics and Laser Symposium (PALS). International Society for Optics and Photonics; p. 991703, (2016).
10- Gurfinkel YI, Sasonko ML, Priezzhev AV. "Digital capillaroscopy as important tool for early diagnostics of arterial hypertension. In: Saratov Fall Meeting 2014: Optical Technologies in Biophysics and Medicine XVI; Laser Physics and Photonics XVI; and Computational Biophysics.", International Society for Optics and Photonics; p. 944804, (2015).
11- Berks M, Dinsdale G, Murray A, Moore T, Manning J, Taylor C, et al. "Automated structure and flow measurement—a promising tool in nailfold capillaroscopy.", Microvasc Res;118:173–7, (2018).
12- Murray AK, Moore TL, Manning JB, Taylor C, Griffiths CE, Herrick AL. "Noninvasive imaging techniques in the assessment of scleroderma spectrum disorders.", Arthritis Care Res.;61(8):1103–11, (2009).
13- Cutolo M, Sulli A, Pizzorni C, Accardo S. "Nailfold videocapillaroscopy assessment of microvascular damage in systemic sclerosis." J Rheumatol;27(1):155–60, (2000).
14- McNutt M. "Reproducibility.", American Association for the Advancement of Science; (2014).
15- Men SJ, Chen C-L, Wei W, Lai T-Y, Song SZ, Wang RK. "Repeatability of vessel density measurement in human skin by OCT-based microangiography.", Skin Res Technol;23(4):607–12, (2017).
16- Cousins CC, Chou JC, Greenstein SH, Brauner SC, Shen LQ, Turalba AV, et al. "Resting nailfold capillary blood flow in primary open-angle glaucoma.", Br J Ophthalmol;103(2):203–7, (2019).
17- Dinsdale G, Moore T, O’Leary N, Berks M, Roberts C, Manning J, et al. "Quantitative outcome measures for systemic sclerosis-related microangiopathy–reliability of image acquisition in nailfold capillaroscopy.", Microvasc Res;113:56–9, (2017).
18- Hofstee HM, Serne EH, Roberts C, Hesselstrand R, Scheja A, Moore TL, et al. "A multicentre study on the reliability of qualitative and quantitative nail-fold videocapillaroscopy assessment.", Rheumatology; 51(4):749–55, (2012).
19- Gasser P. "Capillary blood cell velocity in finger nailfold: Characteristics and reproducibility of the local cold response.", Microvasc Res;40(1):29–35, (1990).
20- Butti P, Intaglietta M, Reimann H, Holliger CH, Bollinger A, Anliker M. "Capillary red blood cell velocity measurements in human nailfold by videodensitometric method.", Microvasc Res;10(2):220–7, (1975).
21- Boss CH, Schneuwly P, Mahler F. "Evaluation and clinical application of the flying spot method in clinical nailfold capillary TV-microscopy.", Int J Microcirc Clin Exp;6(1):15–23,b(1987).
22- Mahler F, Saner H, Boss C, Annaheim M. "Local cold exposure test for capillaroscopic examination of patients with Raynaud’s syndrome.", Microvasc Res;33(3):422–7, (1987).
23- Bollinger A, Butti P, Barras J-P, Trachsler H, Siegenthaler W. "Red blood cell velocity in nailfold capillaries of man measured by a television microscopy technique.", Microvasc Res;7(1):61–72, (1974).
24- Stücker M, Baier V, Reuther T, Hoffmann K, Kellam K, Altmeyer P. "Capillary blood cell velocity in human skin capillaries located perpendicularly to the skin surface: measured by a new laser Doppler anemometer.", Microvasc Res;52(2):188–92, (1996).
25- Wu C-C, Zhang G, Huang T-C, Lin K- P. "Red blood cell velocity measurements of complete capillary in finger nail-fold using optical flow estimation.", Microvasc Res;78(3):319–24, (2009).
Files
IssueVol 10 No 1 (2023) QRcode
SectionOriginal Article(s)
DOI https://doi.org/10.18502/fbt.v10i1.11511
Keywords
Capillary Nailfold Capillaroscopy ImageJ Software Scale-Invariant Feature Transform Tracking Algorithm Kalman Filter

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Pakbin M, Marjaneh Hejazi S, Najafizadeh SR. Quantitative Nail Fold Capillary Blood Flow Using Capillaroscopy System and ImageJ Software in Healthy Individuals. Frontiers Biomed Technol. 2022;10(1):38-46.