Manifold-Guided Multi-Reference Registration for Left-Ventricular Motion Tracking in 3D Echocardiography

Abstract

Purpose: To develop and evaluate a data-driven multi-reference registration strategy for left-ventricular motion tracking in three-dimensional echocardiography, using an intrinsic low-dimensional cardiac trajectory to determine the temporal placement of local registration references.

Materials and Methods: Twelve healthy-volunteer three-dimensional echocardiographic sequences from a publicly available myocardial-tracking benchmark were analyzed. We evaluated six low-dimensional representation methods (PCA, t-SNE, Diffusion Maps, LLE, UMAP, and Isomap) in standard and time-embedded forms using geometric and physiologically relevant criteria. Four reference frames were automatically extracted from the resulting two-dimensional trajectories. Time-embedded PCA and Diffusion Maps produced identical four-reference sets for all analyzed cases and therefore resulted in the same proposed registration configuration. We compared five primary registration configurations—single-reference, sequential frame-to-frame, the proposed manifold-guided multi-reference strategy, a partial B-Spline–Fourier (BSF) implementation, and a Fixed Four-Reference Control—using landmark-based Target Registration Error (TRE) in millimeters. We also evaluated an additional Two-Reference Control (K=2) that retained only the initial ED and principal internal references as a reference-number sensitivity analysis.

Results: For both time-embedded PCA and Diffusion Maps, the automatically selected principal internal reference coincided with the expert-defined end-systolic (ES) frame in 10 of 12 cases and differed by one frame in the remaining two cases. The proposed strategy achieved mean-cycle, ES-frame, and final-frame TRE values of 6.8 ± 1.6 mm, 11.5 ± 2.7 mm, and 4.0 ± 1.2 mm, respectively. For mean-cycle TRE, Strategy 3 showed a median paired reduction of 0.99 mm relative to the Fixed Four-Reference Control [95% CI: 0.94–1.08; |r_rb|=0.91; p_Holm=0.033]. At ES, the corresponding reduction relative to the Fixed Four-Reference Control was 2.000 mm [1.945–2.055; |r_rb|=1.000; p_Holm=0.021]. The K=2 control showed no significant mean-cycle difference from the other configurations, but Strategy 3 reduced ES-frame TRE relative to K=2 by 1.265 mm [0.304–2.445; |r_rb|=0.909; p_Holm=0.029] and final-frame TRE by 1.156 mm [0.483–2.142; |r_rb|=1.000; p_Holm=0.029]. No statistically significant final-frame difference was detected between Strategy 3 and the Fixed Four-Reference Control.

Conclusion: Within this benchmark of 12 healthy-volunteer sequences, the findings support the internal technical feasibility of using the intrinsic low-dimensional organization of a cardiac image sequence for data-driven reference-frame placement within a multi-reference registration architecture. Time-embedded PCA and Diffusion Maps produced identical reference configurations, indicating no additional reference-selection benefit from the more computationally demanding Diffusion Maps representation in the present benchmark. The K=2 sensitivity analysis showed phase-specific effects of removing the intermediate references, while the absence of a significant mean-cycle difference prevents interpretation of K=4 as uniformly superior or empirically optimal. Independent validation in larger and clinically heterogeneous datasets is required before broader generalization.

1. Roth GA, Mensah GA, Johnson CO, et al. Global burden of cardiovascular diseases and risk factors, 1990–2019: update from the GBD 2019 Study. J Am Coll Cardiol. 2020;76:2982–3021.
2. Lang RM, Badano LP, Mor-Avi V, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28(1):1–39.e14.
3. Pellikka PA, Arruda-Olson A, Chaudhry FA, et al. Guidelines for performance, interpretation, and application of stress echocardiography in ischemic heart disease. J Am Soc Echocardiogr. 2020;33(1):1–41.e8.
4. Lang RM, Badano LP, Tsang W, et al. EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography. J Am Soc Echocardiogr. 2012;25(1):3–46.
5. Cerqueira MD, Weissman NJ, Dilsizian V, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105(4):539–542.
6. Ciampi Q, Picano E, Paterni M, et al. Quality control of regional wall motion analysis in stress Echo 2020. Int J Cardiol. 2017;249:479–485.
7. Perperidis A. Postprocessing approaches for the improvement of cardiac ultrasound B-mode images: a review. IEEE Trans Ultrason Ferroelectr Freq Control. 2016;63(3):470–485.
8. American Society of Echocardiography. Recommendations for the Identification and Mitigation of Cardiac Ultrasound Artifacts. J Am Soc Echocardiogr. 2026.
9. Voigt JU, Pedrizzetti G, Lysyansky P, et al. Definitions for a common standard for 2D speckle tracking echocardiography. J Am Soc Echocardiogr. 2015;28(2):183–193.
10. Smiseth OA, Torp H, Opdahl A, Haugaa KH, Urheim S. Myocardial strain imaging: how useful is it in clinical decision making? Eur Heart J. 2016;37(15):1196–1207.
11. Dandel M, Lehmkuhl H, Knosalla C, et al. Strain and strain rate imaging by echocardiography. Int J Cardiol. 2009;132(1):11–24.
12. Crosby J, Amundsen BH, Hergum T, et al. 3-D speckle tracking for assessment of regional left ventricular function. Ultrasound Med Biol. 2009;35(3):458–471.
13. Blessberger H, Binder T. Two-dimensional speckle tracking echocardiography: basic principles. Heart. 2010;96(9):716–722.
14. Seo Y, Ishizu T, Atsumi A, et al. Three-dimensional speckle tracking echocardiography. Circ J. 2014;78(6):1290–1301.
15. Ledesma-Carbayo MJ, Kybic J, Desco M, et al. Spatio-temporal nonrigid registration for ultrasound cardiac motion estimation. IEEE Trans Med Imaging. 2005;24(9):1113–1126.
16. De Craene M, Piella G, Camara O, et al. Temporal diffeomorphic free-form deformation: application to motion and strain estimation from 3D echocardiography. Med Image Anal. 2012;16(2):427–450.
17. Zhang Z, Ashraf M, Sahn DJ, Song X. Temporally diffeomorphic cardiac motion estimation from three-dimensional echocardiography by minimization of intensity consistency error. Med Phys. 2014;41(5):052902.
18. Sundar H, Litt H, Shen D. Estimating myocardial motion by 4D image warping. Pattern Recognit. 2009;42(11):2514–2526.
19. Tascón-Vidarte JD, et al. Accuracy and consistency of intensity-based deformable image registration in 4DCT. PLoS One. 2022;17:e0271064.
20. Wiputra H, Chan WX, Foo YY, et al. Cardiac motion estimation from medical images: a regularisation framework applied to pairwise image registration displacement fields. Sci Rep. 2020;10:18510. doi:10.1038/s41598-020-75525-4.
21. Yang L, Georgescu B, Zheng Y, Meer P, Comaniciu D. 3D ultrasound tracking of the left ventricle using one-step forward prediction and data fusion of collaborative trackers. In: 2008 IEEE Conference on Computer Vision and Pattern Recognition. 2008. doi:10.1109/CVPR.2008.4587518.
22. Rohé MM, Sermesant M, Pennec X. Low-dimensional representation of cardiac motion using Barycentric Subspaces: a new group-wise paradigm for estimation, analysis, and reconstruction. Med Image Anal. 2018;45:1–12. doi:10.1016/j.media.2017.12.008.
23. Wachinger C, Yigitsoy M, Rijkhorst EJ, Navab N. Manifold learning for image-based breathing gating in ultrasound and MRI. Med Image Anal. 2012;16(4):806–818.
24. Gifani P, Behnam H, Shalbaf A, Alizadeh Sani Z. Automatic detection of end-diastole and end-systole from echocardiography images using manifold learning. Physiol Meas. 2010;31(9):1091–1103.
25. Izenman AJ. Introduction to manifold learning. WIREs Comput Stat. 2012;4(5):439–446. doi:10.1002/wics.1222.
26. Lee JA, Verleysen M. Nonlinear Dimensionality Reduction. New York: Springer; 2007.
27. van der Maaten LJP, Postma EO, van den Herik HJ. Dimensionality Reduction: A Comparative Review. Tilburg University Technical Report TiCC-TR 2009-005; 2009.
28. Tobon-Gomez C, De Craene M, McLeod K, et al. Benchmarking framework for myocardial tracking and deformation algorithms: an open access database. Med Image Anal. 2013;17(6):632–648. doi:10.1016/j.media.2013.03.008.
29. Takens F. Detecting strange attractors in turbulence. In: Rand D, Young LS, editors. Dynamical Systems and Turbulence, Warwick 1980. Lecture Notes in Mathematics. Vol 898. Berlin: Springer; 1981. p. 366–381.
30. Jolliffe IT. Principal Component Analysis. 2nd ed. New York: Springer; 2002.
31. van der Maaten L, Hinton G. Visualizing data using t-SNE. J Mach Learn Res. 2008;9:2579–2605.
32. Coifman RR, Lafon S. Diffusion maps. Appl Comput Harmon Anal. 2006;21(1):5–30.
33. Nadler B, Lafon S, Coifman RR, Kevrekidis IG. Diffusion maps, spectral clustering and reaction coordinates of dynamical systems. Appl Comput Harmon Anal. 2006;21(1):113–127.
34. Roweis ST, Saul LK. Nonlinear dimensionality reduction by locally linear embedding. Science. 2000;290(5500):2323–2326.
35. McInnes L, Healy J, Melville J. UMAP: Uniform Manifold Approximation and Projection for Dimension Reduction. arXiv:1802.03426. 2018.
36. Tenenbaum JB, de Silva V, Langford JC. A global geometric framework for nonlinear dimensionality reduction. Science. 2000;290(5500):2319–2323.
37. Venna J, Kaski S. Neighborhood preservation in nonlinear projection methods: an experimental study. In: Artificial Neural Networks — ICANN 2001. Berlin: Springer; 2001. p. 485–491.
38. Lee JA, Verleysen M. Quality assessment of dimensionality reduction: rank-based criteria. Neurocomputing. 2009;72(7–9):1431–1443.
39. Kruskal JB. Multidimensional scaling by optimizing goodness of fit to a nonmetric hypothesis. Psychometrika. 1964;29:1–27.
40. Thirion JP. Image matching as a diffusion process: an analogy with Maxwell's demons. Med Image Anal. 1998;2(3):243–260. doi:10.1016/S1361-8415(98)80022-4.
41. Vercauteren T, Pennec X, Perchant A, Ayache N. Diffeomorphic demons: efficient non-parametric image registration. NeuroImage. 2009;45(1 Suppl): S61–S72. doi:10.1016/j.neuroimage.2008.10.040.
Files
IssueArticles in Press QRcode
SectionOriginal Article(s)
Keywords
3D echocardiography left ventricular motion image registration manifold learning myocardial tracking

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Rashidi M, Jafarnia dabanloo N, Nowshiravan Rahatabad F, Shojaeifard M. Manifold-Guided Multi-Reference Registration for Left-Ventricular Motion Tracking in 3D Echocardiography. Frontiers Biomed Technol. 2026;.