Original Article

Performance Evaluation of a Solid-State Detector Based Handheld Gamma Camera System

Abstract

Purpose: In this work we present a handheld, solid-state detector based gamma camera system. We validated the camera physically and demonstrated the usefulness of the device for intra-operative detection of radiolabeled tissue.
Methods: We measured the intrinsic uniformity, the intrinsic energy resolution, the system uniformity, the spatial resolution without scatter, the system planar uniformity, the detector shielding and the peak deviation according to the NEMA NU1-2001 standard.
Results: The gamma camera can be used for isotopes with an energy range between 50-250 keV. A standard laptop is used to control of the camera and to visualize Preliminary clinical data show that the devices can be used successfully for a number of clinical applications.
Conclusion: The performance evaluation of a novel handheld gamma camera shows good spatial resolution, sensitivity and energy resolution. Due to the small size and weight the portable device can be used for untraoperative acquisitions.

Files
IssueVol 1 No 1 (2014) QRcode
SectionOriginal Article(s)
Keywords
Handheld gamma camera Solid state detector Radiopharmaceutical guided surgery.

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Knoll P, Mirzaei S, Schwenkenbecher K, Barthel T. Performance Evaluation of a Solid-State Detector Based Handheld Gamma Camera System. Frontiers Biomed Technol. 2014;1(1):61-67.