Filters Effect on Image Quality of 177Lu-DOTATATE SPECT Images for Neuro-Endocrine Tumors Scanning
Abstract
Purpose: The quality of the image in SPECT scans depends on the imaging parameters which are determined experimentally in the field of nuclear medicine. Designing a dedicated scanning parameter for 177Lu-DOTATATE SPECT images is required to optimize reconstruction. Therefore, this study aims to evaluate the effect of different filters on image quality for bone SPECT scanning using 177Lu-DOTATATE.
Materials and Methods: The filtered back-projection reconstruction method was used in neuro-endocrine tumor scanning using 177Lu-DOTATATE. Three hours after injection of 177Lu-DOTATATE, SPECT scans from 177Lu-DOTATATE for 30 patients were acquired using a dual-head EvoExel detector system. Seven parameters were considered, including the contrast/noise ratio, injection activity, uptake duration, acquisition time per injection, frame time, measuring time, and type of filters.
Results: In all cases, the application of different filters increased Contrast to Noise Ratio (CNR) (9.1%, 1.8%, 10.9%, 61.8%, 23.6%, 29.1%, and 58.2% for Wiener, Butterworth, Parzen, Metz, Ramp, Shepp-Logan, and Hamming filters, respectively). The percentage of increase in Signal-to-Noise Ratio (SNR) was 3.3%, 1.7%, -24%, 21.7%, 9.8%, 11.9%, and 20.6%.
Conclusion: Based on the quantitative analysis of the results, the application of the Metz filter (power 2) and Hamming filter (with 0.27, 0.47, and 0.67 cut-off frequencies) on SPECT scans of neuro-endocrine tumors is recommended because of their capacity to provide high-quality images.
Files | ||
Issue | Articles in Press | |
Section | Original Article(s) | |
Keywords | ||
Nuclear Medicine Image Reconstruction Filter Design Image Quality Single Photon Emission Computed Tomography Imaging |